The Weston cell or Weston standard cell is a wet cell that produces a highly stable voltage suitable as a laboratory standard for calibration of . Invented by Edward Weston in 1893, it was adopted as the International Standard for EMF from 1911 until superseded by the Josephson voltage standard in 1990.
As shown in the illustration, the cell is set up in an H-shaped glass vessel with the cadmium amalgam in one leg and the pure mercury in the other. Electrical connections to the cadmium amalgam and the mercury are made by platinum wires fused through the lower ends of the legs.
Reference cells must be applied in such a way that no current is drawn from them.
One of the great advantages of the Weston normal cell is its small change of electromotive force with change of temperature. At any temperature between and ,
The temperature coefficient can be reduced by shifting to an unsaturated design, the predominant type today. However, an unsaturated cell's output decreases by some 80 microvolts per year, which is compensated by periodic calibration against a saturated cell.
|
|